You don't have or don't remember the password! Click Here
Editor-in-chief
Maria Stella Graziani
Deputy Director
Martina Zaninotto
Associate Editors
Ferruccio Ceriotti
Davide Giavarina
Bruna Lo Sasso
Giampaolo Merlini
Martina Montagnana
Andrea Mosca
Paola Pezzati
Rossella Tomaiuolo
Matteo Vidali
EIC Assistant
Francesco Busardò
International Advisory Board
Khosrow Adeli Canada
Sergio Bernardini Italy
Marcello Ciaccio Italy
Eleftherios Diamandis Canada
Philippe Gillery France
Kjell Grankvist Sweden
Hans Jacobs The Netherlands
Eric Kilpatrick UK
Magdalena Krintus Poland
Giuseppe Lippi Italy
Mario Plebani Italy
Sverre Sandberg Norway
Ana-Maria Simundic Croatia
Tommaso Trenti Italy
Cas Weykamp The Netherlands
Maria Willrich USA
Paul Yip Canada
Publisher
Biomedia srl
Via L. Temolo 4, 20126 Milano
Responsible Editor
Giuseppe Agosta
Editorial Secretary
Chiara Riva
Biomedia srl
Via L. Temolo 4, 20126 Milano
Tel. 0245498282
email: biochimica.clinica@sibioc.it
--------------------
ISSN print: 0393 – 0564
ISSN digital: 0392- 7091
BC: Articoli scritti da C. Passarello
Valutazione multicentrica dell’analizzatore Tosoh G8 per la misura dell’emoglobina A2 e dell’emoglobina F
Multicenter evaluation of the Tosoh G8 analyzer for determination of hemoglobin (Hb) A2 and F
<p>The analytical performance of the new Tosoh automated analyzer HLC-723 G8 (-thalassemia analysis mode) to determine hemoglobin variants and to measure HbA2 and HbF in human blood was evaluated in three Italian centres. The within- and between-run imprecision for HbA2 were good, with CV between 0.2% and 1.8% and between 0.9% and 5.4%, respectively. The CV for HbF was between 0.4% and 9.8% (within-run) and beetwen 0.8% and 13.1% (between-run). The comparability of HbA2 measurements between different centres was excellent (r=0.99), but a significant bias in comparison with the previous version of the instrument was noted. Experiments to test the HbA2 stability in blood confirmed that blood samples stored at -80 °C were stable for at least 4 months and that storage at -20 °C is not recommended. In conclusion, the Tosoh G8 analyser was found reliable and robust and, therefore, suitable for the measurement of HbA2 and HbF in human blood.</p>
Gentile utente, le ricordiamo che l'accesso alla rivista è riservato ai soci SIBioC in regola con la quota associativa.
Per accedere alla pagina richiesta è necessario inserire il suo nome utente e la password nella sezione "AREA SOCI".
A disposizione per ulteriori informazioni.
La Segreteria SIBioC
Dear User, please note that access to the Journal is for SIBioC members who have renewed their membership fee.
To access the requested page you need to enter username and password in "Area Soci".
At your disposal for further information.
SIBioC Secretariat