You don't have or don't remember the password! Click Here
Editor-in-chief
Maria Stella Graziani
Deputy Director
Martina Zaninotto
Associate Editors
Ferruccio Ceriotti
Davide Giavarina
Bruna Lo Sasso
Giampaolo Merlini
Martina Montagnana
Andrea Mosca
Paola Pezzati
Rossella Tomaiuolo
Matteo Vidali
EIC Assistant
Francesco Busardò
International Advisory Board
Khosrow Adeli Canada
Sergio Bernardini Italy
Marcello Ciaccio Italy
Eleftherios Diamandis Canada
Philippe Gillery France
Kjell Grankvist Sweden
Hans Jacobs The Netherlands
Eric Kilpatrick UK
Magdalena Krintus Poland
Giuseppe Lippi Italy
Mario Plebani Italy
Sverre Sandberg Norway
Ana-Maria Simundic Croatia
Tommaso Trenti Italy
Cas Weykamp The Netherlands
Maria Willrich USA
Paul Yip Canada
Publisher
Biomedia srl
Via L. Temolo 4, 20126 Milano
Responsible Editor
Giuseppe Agosta
Editorial Secretary
Chiara Riva
Biomedia srl
Via L. Temolo 4, 20126 Milano
Tel. 0245498282
email: biochimica.clinica@sibioc.it
--------------------
ISSN print: 0393 – 0564
ISSN digital: 0392- 7091
BC: Articoli scritti da R.M. Paragliola
Valori falsamente elevati di triodotironina libera in una paziente affetta da tiroidite cronica e gozzo multinodulare
Falsely increased free triiodothyronine values in a woman affected by thyroiditis and multinodular goiter
<p>We describe a case of a 63-years old woman affected by Hashimoto’s thyroiditis and multinodular goiter. Her laboratory results showed elevated free triiodothyronine (FT3) concentrations (18.5 ng/L), with free thyroxine (FT4) and thyrotropin (TSH) within the physiologic range. On the basis of these results, she started methimazole therapy, with persistence of inappropriately elevated FT3 concentrations. The therapy was thus stopped and blood examination was repeated after one month in our laboratory: concentrations of thyroid tests were within the physiologic range, including FT3 (2.8 ng/L). The difference between this result and that previously obtained raised the suspicion of the presence of an interference in the first result. In fact, in our laboratory a competitive electrochemiluminescence immunoassay with labeled antibody is used, while the first laboratory employed a competitive chemiluminescence immunoassay with labeled analogue, which has a more important risk of interference. After treating sample by polyethylene glycol, FT3 resulted indeed normal also by the immunoassay used by the first laboratory.</p>
Biochimica Clinica ; 38(1) 70-72 Casi clinici - Case report
Gentile utente, le ricordiamo che l'accesso alla rivista è riservato ai soci SIBioC in regola con la quota associativa.
Per accedere alla pagina richiesta è necessario inserire il suo nome utente e la password nella sezione "AREA SOCI".
A disposizione per ulteriori informazioni.
La Segreteria SIBioC
Dear User, please note that access to the Journal is for SIBioC members who have renewed their membership fee.
To access the requested page you need to enter username and password in "Area Soci".
At your disposal for further information.
SIBioC Secretariat